Add to Cart
Equipment Development Overview
This device is suitable for leak detection of Prismastic cell batteries. Manually picking and discharging materials, automatically detecting leaks and providing detection results, no alarm if OK, and an alarm if NG.
Technical parameters and configuration characteristics
1. Number of detectable workpieces 1
2. Minimum detectable leakage rate 3 × 10-13Pa · m3/s
3. Leakage rate display range: 1 × 10-1------1 × 10-13Pa · m3/s
4. Start time ≤ 100s
5. Response time ≤ 0.3 S
6. The maximum pressure at the leak detection port is 2000Pa
7. Power requirements: AC220V ± 10%/50Hz
8. Leakage rate units: Pa · m3/s, ppm, mbar · l/s, atm.cc/s
9. Display interface with a large 7.0-inch touch screen display, clear and easy to understand operation interface
10. Unqualified leakage rate setting, alarm volume setting, etc., equipped with automatic leak detection alarm, providing sound and visual prompts
11. With serial RS232 level output
12. Large capacity memory, capable of storing 10000 sets of test results
13. USB data output, which can directly write detection data to a USB drive
14. Rich input and output contacts
15. With a complete product chain and a variety of optional accessory combinations, it can meet the testing needs of different products
16. Main components imported from Europe and America: original imported molecular pump, German Leibao mechanical pump, German Baoti electromagnetic valve, American imported amplifier